Lambda
LS
Stand-Alone Xenon Arc Lamp and Power Supply
|

Shown with optional
liquid light guide
|
 |
|
|
| S
P E C S > L
A M B D A L S |
Output
Range
Standard bulb 340nm to IR
(optional full-spectrum bulb) 200nm to IR |
Lamp
Type
175 or 300 Watt xenon (pre-aligned to produce collimated output) |
Radiant
Output
25 Watts (175W lamp) (broadband, full beam) |
Lamp
Life
1000 hours (Bulb warranted for 500 hours. Longer life depends on application.
Expected life is 1000 hours.) |
Power
Consumption
175 Watts |
Dimensions
10.5in x 9.5in x 10in
26.7cm x 24.1cm x 25.4cm |
Weight
10.5lbs
4.8kg |
Electrical
115/230 Volts
50/60 Hertz power line |
|
|
|
 |
The Lambda LS is a stand-alone illuminator
consisting of a xenon-arc lamp, lamp housing, cold mirror and power supply
in a single enclosure. The Lambda LS is designed to be used with
a liquid light guide which transmits remarkably flat, intense, illumination
to the optical train of the users microscope or other instrumentation.
The lamps cabinet accommodates a standard Sutter Instrument filter
wheel that slides easily in and out of a slot in the light path. If desired,
a second filterwheel can be mounted on the outside of the cabinet. When
used with appropriate adapters, the light guide output is compatible with
most common microscope systems.
| F
E A T U R E S > L A M B D A L S |
| Xenon
lamps provide light levels which exceed those of standard microscope
fluorescence lamps. |
| Equipped
with a cold mirror to eliminate IR heating of down stream optical
components. |
| Compact
stand-alone lamp housing-power supply enclosure. |
| Pre-aligned
bulb eliminates common focusing problems. |
| Integrated
hour meter for convenient monitoring of lamp life. |
| Modular
construction allows use of optional liquid light guide for flexible
direction of light output. |
| Easily
accommodates Sutter Instrument 25mm filter wheels within the body
of the lamp. |
Can
be coupled via a liquid light guide to many standard microscopes (Nikon,
Zeiss, Leica and Olympus). Microscope coupling requires special
adapters (not included); please see the adapter section in this catalog. |
Unlike the arc lamps used with most fluorescence microscopes,
the 175W xenon bulb is pre-aligned using a parabolic mirror and does not
require alignment, focusing or collimation. In the standard configuration,
the Lambda LS bulb is capable of producing light output from 340nm
to a cutoff of 700nm determined by the cold mirror. An optional enhanced
UV bulb produces output much lower into the UV (cut off near 200nm). As
with any UV generating light source, the optional bulb generates significant
quantities of ozone and must be used in an adequately ventilated environment.
The Lambda LS utilizes a compact design, which places power supply,
lamp house, arc lamp and cold mirror in a single enclosure. This system
eliminates a common failure associated with standard arc lamp designs;
when using a remote power supply aging may lead to a decreased ability
to light the lamp due to loss of the insulating characteristics of the
lengthy high-tension line. As with all our equipment, the power supply
has been designed to minimize electrical noise that can be picked up by
physiological recording equipment.

|